Stray radiation dose reduction using a backscatter shield and a rectangular collimator for a handheld intraoral X-ray
Yusei Otaka1,2, Yasuo Harata1, Maki Izawa1
1Division of Dental Radiology, Department of Diagnostic and Therapeutic Sciences, Meikai University School of Dentistry, 1-1 Keyakidai, Sakado, Saitama 350-0283, Japan.
Abstract:
The objective of this study was to study the dose reduction effect on occupational exposure of operators and public exposure other than operators when a backscatter shield and a rectangular collimator are used in conjunction with a handheld intraoral X-ray unit. The occupational exposure was reduced to 40% when the backscatter shield was attached to the cone-tip, to 13% when the rectangular collimator was attached and to 7.7% when the backscatter shield and rectangular collimator were used together. On the other hand, the public exposure was reduced to 20% when the rectangular collimator was attached, but the backscatter shield was not effective in reducing the public exposure. Attaching a backscatter shield is effective in reducing the occupational exposure, and a rectangular aperture is effective in reducing the occupational exposure, as well as the public exposure.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
X-ray Imaging
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...
Biological Effects of Radiation