Validation of an AI-based algorithm for measurement of the thoracic aortic diameter in low-dose chest CT
I Iris Hamelink1, E Erik Jan de Heide1, G J Gert Jan Pelgrim1
1Department of Radiology, University of Groningen, University Medical Center of Groningen, 9713GZ Groningen, The Netherlands.
Objectives:
To evaluate the performance of artificial intelligence (AI) software for automatic thoracic aortic diameter assessment in a heterogeneous cohort with low-dose, non-contrast chest computed tomography (CT).
Materials And Methods:
Participants of the Imaging in Lifelines (ImaLife) study who underwent low-dose, non-contrast chest CT (August 2017-May 2022) were included using random samples of 80 participants <50y, ≥80y, and with thoracic aortic diameter ≥40 mm. AI-based aortic diameters at eight guideline compliant positions were compared with manual measurements. In 90 examinations (30 per group) diameters were reassessed for intra- and inter-reader variability, which was compared to discrepancy of the AI system using Bland-Altman analysis, paired samples t-testing and linear mixed models.
Results:
We analyzed 240 participants (63 ± 16 years; 50 % men). AI evaluation failed in 11 cases due to incorrect segmentation (4.6 %), leaving 229 cases for analysis. No difference was found in aortic diameter between manual and automatic measurements (32.7 ± 6.4 mm vs 32.7 ± 6.0 mm, p = 0.70). Bland-Altman analysis yielded no systematic bias and a repeatability coefficient of 4.0 mm for AI. Mean discrepancy of AI (1.3 ± 1.6 mm) was comparable to inter-reader variability (1.4 ± 1.4 mm); only at the proximal aortic arch showed AI higher discrepancy (2.0 ± 1.8 mm vs 0.9 ± 0.9 mm, p < 0.001). No difference between AI discrepancy and inter-reader variability was found for any subgroup (all: p > 0.05).
Conclusion:
The AI software can accurately measure thoracic aortic diameters, with discrepancy to a human reader similar to inter-reader variability in a range from normal to dilated aortas.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:08Ultrasound Imaging of the Thoracic and Abdominal Aorta in Mice to Determine Aneurysm Dimensions
Published on: March 8, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT