A noise-controlling method by hybrid current-stimulation and voltage-measurement for electrical impedance tomography
Zengfeng Gao1, Panji Nursetia Darma2, Bo Sun3
1Division of Fundamental Engineering, Graduate School of Science and Engineering, Chiba University, Chiba 263-8522, Japan.
None:
Image reconstruction in electrical impedance tomography (EIT) is a typical ill-posed inverse problem, from which the stability of conductivity reconstruction affects the reliability of physiological parameters evaluation. In order to improve the stability, the effect of boundary voltage noise on conductivity reconstruction should be controlled. A noise-controlling method based on hybrid current-stimulation and voltage-measurement for EIT (HCSVM-EIT) is proposed for stable conductivity reconstruction. In HCSVM-EIT, the boundary voltage is measured by one current-stimulation and voltage-measurement pattern (high-SNRpattern) with a higher signal-to-noise ratio (SNR); the sensitivity matrix is calculated by another current-stimulation and voltage-measurement pattern (low-condpattern) with a lower condition number; the boundary voltage is then transformed from thehigh-SNRpattern into thelow-condpattern by multiplying by an optimized transformation matrix for image reconstruction. The stability of conductivity reconstruction is improved by combining the advantages of thehigh-SNRpattern for boundary voltage measurement and thelow-condpattern for sensitivity matrix calculation. The simulation results show that the HCSVM-EIT increases the correlation coefficient (CC) of conductivity reconstruction. The experiment results show that theCCof conductivity reconstruction of the human lower limb is increased from 0.3424 to 0.5580 by 62.97% compared to the quasi-adjacent pattern, and from 0.4942 to 0.5580 by 12.91% compared to the adjacent pattern. In conclusion, the stable conductivity reconstruction with higherCCin HCSVM-EIT improves the reliability of physiological parameters evaluation for disease detection.
Related Concept Videos
Interfacial Electrochemical Methods: Overview
Electronic Distance Measuring Instruments