Are mammography image acquisition factors, compression pressure and paddle tilt, associated with breast cancer
Sue M Hudson1, Louise S Wilkinson2, Bianca L De Stavola3
1Department of Non-Communicable Disease Epidemiology, London School of Hygiene and Tropical Medicine, London, United Kingdom.
Objectives:
To assess the associations between objectively measured mammographic compression pressure and paddle tilt and breast cancer (BC) detected at the same ("contemporaneous") screen, subsequent screens, or in-between screens (interval cancers).
Methods:
Automated pressure and paddle tilt estimates were derived for 80,495 mammographic examinations in a UK population-based screening programme. Adjusted logistic regression models were fitted to estimate the associations of compression parameters with BC detected at contemporaneous screen (777 cases).Nested case-control designs were used to estimate associations of pressure and tilt with: (a) interval cancer (148 cases/625 age-matched controls) and (b) subsequent screen-detected cancer (344/1436), via conditional logistic regression.
Results:
Compression pressure was negatively associated with odds of BC at contemporaneous screen (odds ratio (OR) for top versus bottom third of the pressure distribution: 0.74; 95% CI 0.60, 0.92; P-for-linear-trend (Pt) = 0.007). There was weak evidence that moderate pressure at screening was associated with lower odds of interval cancer (OR for middle versus bottom third: 0.63; 95% CI 0.38, 1.05; p = 0.079), but no association was found between pressure and the odds of BC at subsequent screen. There was no evidence that paddle tilt was associated with the odds of contemporaneous, subsequent screen or interval cancer detection.
Conclusions:
Findings are consistent with compression pressure, but not paddle tilt, affecting the performance of mammographic screening by interfering with its ability to detect cancers.
Advances In Knowledge:
Inadequate or excessive compression pressure at screening may contribute to a reduced ability to detect cancers, resulting in a greater number of interval cancer cases.
More Related Videos
15:48Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
06:50Troubleshooting FoCUS Image Acquisition: Patient Positioning, Transducer Manipulation, and Image Optimization
Published on: March 3, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...