Analyzing the Interaction of Fluorescent-Labeled Proteins with Artificial Phospholipid Microvesicles using Quantitative Flow Cytometry
Published on: April 6, 2022
2.5K
Kinetic titration method in flow cytometry for quantification of cell receptors
Irina V Khalo1, Andrei A Karpenko2, Viktoriya S Kozyreva2
1Voevodsky Institute of Chemical Kinetics and Combustion SB RAS, Institutskaya 3, 630090 Novosibirsk, Russia.
Journal of Immunological Methods
|September 4, 2023
Summary
This study introduces a new flow cytometry method for quantifying cell receptors using reaction kinetics, eliminating the need for calibration beads. The method revealed reduced CD14 receptors on leukocytes in atherosclerosis patients compared to healthy donors.
Area of Science:
- Immunology
- Biophysics
- Analytical Chemistry
Background:
- Accurate quantification of cell surface receptors is crucial for understanding cellular function and disease.
- Traditional methods for receptor quantification using fluorescence flow cytometry have limitations, including reliance on calibration beads and potential inaccuracies due to reaction kinetics.
Purpose of the Study:
- To develop and validate a novel, calibration-bead-free method for quantitative determination of cell receptors using fluorescence flow cytometry.
- To account for reaction kinetics in receptor-ligand binding for more accurate cell surface receptor quantification.
Main Methods:
- A new method based on reaction kinetics was proposed, involving samples with constant cell concentration and varying initial fluorescently labeled ligand concentrations.
- Each sample was measured by flow cytometry when the product of incubation time and initial ligand concentration was constant across all samples.
- The method was applied to quantify CD8 and CD14 receptors on human peripheral blood leukocytes.
Main Results:
- The proposed method successfully quantified CD8 and CD14 receptors without using calibration beads.
- A significant decrease (two-fold) in the mean number of CD14 receptors was observed on monocytes and granulocytes of atherosclerosis patients compared to healthy donors.
- No significant difference in mean CD8 expression was found on leukocytes between the patient and donor groups.
Conclusions:
- The developed kinetic method offers an alternative approach for accurate cell receptor quantification by flow cytometry, overcoming limitations of existing methods.
- The findings suggest a potential role for reduced CD14 expression in the pathophysiology of atherosclerosis.
- The method provides a valuable tool for immunological research and clinical diagnostics.