Electromagnetic interference from automobile passive keyless entry in cardiovascular implantable electronic devices
Narawudt Prasertwitayakij1, Taradol Komolmis2, Siriluck Gunaparn1
1Division of Cardiology, Department of Internal Medicine, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand.
Cardiovascular implantable electronic device (CIED) patients can safely use automobile passive keyless entry (PKE) systems. This study found no electromagnetic interference (EMI) from PKE systems, indicating no risk to CIED function.
Area of Science:
- Biomedical Engineering
- Cardiology
- Electromagnetics
Background:
- Automobile passive keyless entry (PKE) systems may generate electromagnetic interference (EMI).
- Cardiovascular implantable electronic devices (CIEDs) are sensitive to EMI.
- The potential for EMI from PKE systems in CIED patients requires investigation.
Purpose of the Study:
- To determine the incidence and significance of EMI from automobile PKE systems in patients with CIEDs.
- To assess the safety of PKE system use for individuals with CIEDs.
Main Methods:
- A single-center, cross-sectional study involving 102 CIED patients.
- Patients operated PKE systems in two automobiles while monitored for EMI and arrhythmias via CIED interrogation and ECG event recorders.
- Spectrum analyzers were used to identify PKE system operating frequencies.
Main Results:
- No incidence of EMI or arrhythmias was detected in CIED patients during PKE system activation.
- The working frequency bandwidth of the PKE systems could not be identified due to low signal strength.
- 102 patients with various CIEDs (defibrillators, pacemakers, CRT devices) participated.
Conclusions:
- Automobile PKE systems emit very low-power signals.
- Under normal conditions, CIED patients can safely use PKE systems without risk of EMI.
- Key fob proximity to the CIED pulse generator does not pose a safety risk.