Related Experiment Video
Updated: Jul 17, 2025

Thermal Measurement Techniques in Analytical Microfluidic Devices
Published on: June 3, 2015
Verifying measurements on Siemens Atellica® instruments using clinically acceptable analytical performance
Emmi Rotgers1, Tea Lamberg1, Tero Pihlajamaa1
1Department of Clinical Chemistry, University of Helsinki, and HUS Diagnostic Center, Helsinki and Uusimaa Hospital District, Helsinki, Finland.
Abstract:
Measurements on clinical chemistry analysers must be verified to demonstrate applicability to their intended clinical use. We verified the performance of measurements on the Siemens Atellica® Solution chemistry analysers against the clinically acceptable analytical performance specifications, CAAPS, including the component of intra-individual biological variation, CVI. The relative standard uncertainty of measurement, i.e. analytical variation, CVA, was estimated for six example measurands, haemoglobin A1c in whole blood (B-HbA1c), albumin in urine (U-Alb), and the following measurands in plasma: sodium (P-Na), pancreatic amylase (P-AmylP), low-density lipoprotein cholesterol (P-LDL-C), and creatinine (P-Crea). Experimental CVA was calculated from single-instrument imprecision using control samples, variation between measurements on parallel instruments, and estimation of bias with pooled patient specimens. Each obtained CVA was compared with previously developed CAAPS. The calculated CVA was 1.4% for B-HbA1c (CAAPS 1.9% for single diagnostic testing, CAAPS 2.0% for monitoring after duplicate tests; IFCC units), 10.9% for U-Alb (CAAPS 44.9%), 1.2% for P-Na (CAAPS 0.6%, after triplicate testing 1.5%), 8.2% for P-AmylP (CAAPS 22.9%). The CVA was 4.9% for P-LDL-C (CAAPS for cardiovascular risk stratification 4.9% after four replicates), and 4.2% for P-Crea (CAAPS 8.0%). Three of the six measurands fulfilled the estimated clinical need. Results from P-Na measurements indicate a general need for improving the P-Na assays for emergency patients. It is necessary to consider CVI when creating diagnostic targets for laboratory tests, as emphasised by the CAAPS estimates of B-HbA1c and P-LDL-C.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Data Validation
Key parameters for method validation include:
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Quality Assurance

