[Method for Measuring Noise Power Spectrum Depending on X-ray Tube Angle in Spiral Orbit of Helical CT Scan]
Yuki Ohsugi1,2, Akihiro Narita1, Masaki Ohkubo1
1Department of Radiological Technology, Graduate School of Health Sciences, Niigata University.
Purpose:
The noise power spectrum (NPS) in computed tomography (CT) images potentially varies with the X-ray tube angle in a spiral orbit of the helical scan. The purpose of this study was to propose a method for measuring the NPS for each angle of the X-ray tube.
Methods:
Images of the water phantom were acquired using a helical scan. As a conventional method, we measured the two-dimensional (2D) NPS from each image and averaged them; the obtained 2D-NPS was referred to as NPSconventional. In the proposed method, we made the X-ray tube angle θ (0°≤θ<360°) to correspond to the image according to each slice position of the images that located within the travel distance of the CT scan table per 360° rotation of the X-ray tube. We obtained the 2D-NPS from each image and assigned the θ (0°, 30°, 60°, 90°, 120°, 150°, 180°); the obtained 2D-NPS was referred to as NPSsθ. The NPSsθ was compared to the NPSconventional. Also, we investigated the dependency of the NPSsθ on the θ.
Results:
The NPSconventional was found to be isotropic, and in contrast, the NPSsθ was anisotropic. The NPSsθ showed a continuously rotational change while increasing the θ. There was an excellent correlation (R2>0.999) between the rotation angle of NPSθ and the θ.
Conclusion:
The proposed method was demonstrated to be effective for evaluating anisotropic noise characteristics depending on the X-ray tube angle.