Abstract:
A spray on, copolymer acrylic dressing (Op-Site) was used to limit the skin damage caused by a transcutaneous oxygen electrode and its adhesive ring. Two identical electrodes were applied to the abdominal skin of 10 preterm infants, one on untreated skin, the other after application of Op-Site. It was found that Op-Site prevented the epidermal damage (as measured by transepidermal water loss) that occurs when the adhesive ring is removed from untreated skin. It did not interfere with transcutaneous oxygen measurements; absolute values and response times were unchanged. Op-Site is therefore useful in preventing the skin trauma that occurs when transcutaneous oxygen monitoring is being performed in preterm infants below 30 weeks' gestation in the first week of life. Care must be taken, however, to prevent a build up of Op-Site--it should be applied as a single layer, allowed to dry, and removed after use.
More Related Videos
10:05Stimulation of Stem Cell Niches and Tissue Regeneration in Mouse Skin by Switchable Protoporphyrin IX-Dependent Photogeneration of Reactive Oxygen Species In Situ
Published on: May 8, 2020
07:22A Standardized Procedure of Dressing Management for Toxic Epidermal Necrolysis
Published on: March 14, 2025
Related Concept Videos
Burn Injuries
The damage results in the death of skin cells, which can lead to a massive loss of fluid. Dehydration, electrolyte imbalance, and renal and circulatory failure follow, which can be fatal. Burn patients are treated with intravenous fluids to offset...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: