Impact of age on measurement variability for axial length in myopic children

Rebecca M Dang1, Kathleen Watt1, Pauline Kang1

  • 1School of Optometry and Vision Science, University of New South Wales, Sydney, Australia.

PubMed

Insights

Axial length measurement variability in myopic children was not affected by age for either the IOLMaster 700 or IOLMaster 500. The IOLMaster 700 provided more consistent axial length readings than the IOLMaster 500.

Area of Science:

  • Ophthalmology
  • Optometry
  • Biomedical Engineering

Background:

  • Axial length is crucial for managing progressive myopia.
  • Younger children's fixation difficulties can increase measurement variability.
  • This variability may lead to inaccurate myopia progression assessment and management.

Purpose of the Study:

  • To evaluate the impact of patient age on axial length measurement variability.
  • To compare the IOLMaster 700 and IOLMaster 500 in myopic children.
  • To determine if age influences measurement consistency.

Main Methods:

  • Retrospective chart review at a university optometry clinic.
  • Collected five axial length measurements per patient using IOLMaster 700 and IOLMaster 500.
  • Calculated within-subject standard deviation and R-squared to assess age-related variability.

Main Results:

  • No significant effect of patient age on measurement variability for either device.
  • IOLMaster 700 measurements were longer (difference -0.02 ± 0.02 mm, p < 0.001).
  • IOLMaster 700 exhibited significantly lower within-subject variability than IOLMaster 500 (p < 0.001).

Conclusions:

  • Axial length measurement variability is independent of age in myopic children.
  • IOLMaster 700 offers superior consistency compared to IOLMaster 500.
  • Clinicians should recognize device differences and avoid interchangeable use, especially in myopia management.