Don't hold PAT: watch for and correct oximetry artifact
David T Plante1,2, Meredith E Rumble1,2
1Department of Psychiatry, University of Wisconsin-Madison, Madison, Wisconsin.
Abstract:
This study evaluated the accuracy of the algorithmic oxygen saturation (SpO2) nadir detection of WatchPAT (Zoll/Itamar, Caesarea, Israel) compared with visual inspection in a real-world setting. SpO2 tracings for 209 consecutive adult WatchPAT recordings were reviewed for SpO2 artifact, with erroneous SpO2 data removed manually. Error rates for SpO2 minima were determined across all studies, and relationships between correct and erroneous studies examined. The overall error rate for SpO2 nadir was 22.5%. Erroneous studies had overall less time spent at SpO2 ≤ 88%, higher true SpO2 nadir, lower mean body mass index, and greater artifact time; however, these variables were not associated with the magnitude of discrepancy between manual and algorithmically derived SpO2 minima. These data demonstrate that SpO2 nadir determined by WatchPAT algorithms should not be considered universally accurate. Like other home sleep apnea tests, visual inspection and manual correction of the study data are often required to derive accurate clinical results.
Citation:
Plante DT, Rumble ME. Don't hold PAT: watch for and correct oximetry artifact. J Clin Sleep Med. 2023;19(12):2113-2116.
More Related Videos
06:11Employing the Forced Oscillation Technique for the Assessment of Respiratory Mechanics in Adults
Published on: February 9, 2022
10:23Equipment Setup and Artifact Removal for Simultaneous Electroencephalogram and Functional Magnetic Resonance Imaging for Clinical Review in Epilepsy
Published on: June 23, 2023
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Errors occurring during blood pressure monitoring
Several factors...