Noise Reduction in Photoplethysmography Signals Using a Convolutional Denoising Autoencoder With Unconventional

Abstract

Related Concept Videos

Downsampling01:20

Downsampling

When considering a sampled sequence with zero values between sampling instants, one can replace it by taking every N-th value of the sequence. At these integer multiples of N, the original and sampled sequences coincide. This process, known as decimation, involves extracting every N-th sample from a sequence, thereby creating a more efficient sequence.
The Fourier transform of the decimated sequence reveals a combination of scaled and shifted versions of the original spectrum. This...
182
Deconvolution01:20

Deconvolution

Deconvolution, also known as inverse filtering, is the process of extracting the impulse response from known input and output signals. This technique is vital in scenarios where the system's characteristics are unknown, and they must be inferred from the observable signals.
Deconvolution involves several mathematical techniques to derive the impulse response. One common approach is polynomial division. In this method, the input and output sequences are treated as coefficients of...
186
Upsampling01:22

Upsampling

Managing signal sampling rates is essential in digital signal processing to maintain signal integrity. A decimated signal, characterized by a reduced frequency range due to its lower sampling rate, can be upsampled by inserting zeros between each sample. This upsampling process expands the original spectrum and introduces repeated spectral replicas at intervals dictated by the new Nyquist frequency. To refine this zero-inserted sequence, it is passed through a lowpass filter with a cutoff...
261
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
1.0K
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
6.4K
Pulse Oximetry01:24

Pulse Oximetry

Pulse oximetry, or SpO2, is a non-invasive method for continuously monitoring arterial oxygen saturation (SaO2). This procedure involves attaching a probe or sensor to the patient's fingertip, forehead, earlobe, or nose bridge. The sensor works by detecting changes in oxygen saturation levels through light signals generated by the oximeter and reflected by the pulsing blood under the probe.
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
357