Evaluation of Artifact Appearance and Burden in Pediatric Brain Tumor MR Imaging with Compressed Sensing in
Rieke Lisa Meister1,2, Michael Groth3, Shuo Zhang4
1Department of Diagnostic and Interventional Radiology and Nuclear Medicine, Section of Pediatric Radiology, University Medical Center Hamburg-Eppendorf, 20251 Hamburg, Germany.
Abstract:
Clinical magnetic resonance imaging (MRI) aims for the highest possible image quality, while balancing the need for acceptable examination time, reasonable signal-to-noise ratio (SNR), and lowest artifact burden. With a recently introduced imaging acceleration technique, compressed sensing, the acquisition speed and image quality of pediatric brain tumor exams can be improved. However, little attention has been paid to its impact on method-related artifacts in pediatric brain MRI. This study assessed the overall artifact burden and artifact appearances in a standardized pediatric brain tumor MRI by comparing conventional parallel imaging acceleration with compressed sensing. This showed that compressed sensing resulted in fewer physiological artifacts in the FLAIR sequence, and a reduction in technical artifacts in the 3D T1 TFE sequences. Only a slight difference was noted in the T2 TSE sequence. A relatively new range of artifacts, which are likely technique-related, was noted in the 3D T1 TFE sequences. In conclusion, by equipping a basic pediatric brain tumor protocol for 3T MRI with compressed sensing, the overall burden of common artifacts can be reduced. However, attention should be paid to novel compressed-sensing-specific artifacts.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Magnetic Resonance Imaging