Related Experiment Video
Updated: Jul 16, 2025

Sensitivity Enhancement of Soft Capacitive Pressure Sensors Using a Solvent Evaporation-Based Porosity Control Technique
Published on: March 24, 2023
An Evaluation of Parylene Thin Films to Prevent Encrustation for a Urinary Bladder Pressure MEMS Sensor System
Sébastien Buchwalder1,2, Mario Hersberger1, Henrike Rebl3
1School of Biomedical and Precision Engineering, University of Bern, Güterstrasse 24/26, 3010 Bern, Switzerland.
Abstract:
Recent developments in urological implants have focused on preventive strategies to mitigate encrustation and biofilm formation. Parylene, a conformal, pinhole-free polymer coating, has gained attention due to its high biocompatibility and chemical resistance, excellent barrier properties, and low friction coefficient. This study aims to evaluate the effectiveness of parylene C in comparison to a parylene VT4 grade coating in preventing encrustation on a urinary bladder pressure MEMS sensor system. Additionally, silicon oxide (SiOx) applied as a finish coating was investigated for further improvements. An in vitro encrustation system mimicking natural urine flow was used to quantify the formation of urinary stones. These stones were subsequently analyzed using Fourier transform infrared spectrometry (FTIR). Encrustation results were then discussed in relation to coating surface chemical properties. Parylene C and VT4 grades demonstrated a very low encrustation mass, making them attractive options for encrustation prevention. The best performance was achieved after the addition of a hydrophilic SiOx finish coating on parylene VT4 grade. Parylene-based encapsulation proved to be an outstanding solution to prevent encrustation for urological implants.