Related Experiment Video
Updated: Jul 4, 2026

Deep Proteome Profiling by Isobaric Labeling, Extensive Liquid Chromatography, Mass Spectrometry, and Software-assisted Quantification
Published on: November 15, 2017
Quantification of human serum albumin by combining chymotrypsin/trypsin digestion coupled with LC-MS/MS technique
Meiyun Shi1, Xujian Duan2, Xinyue Zheng2
1School of Life and Pharmaceutical Sciences, Dalian University of Technology, Panjin, 1242(21), PR China; Aim Honesty Biopharmaceutical Co. LTD, Dalian, 116600, PR China.
Abstract:
The quantification of albumin is important in clinical medicine because the concentration of albumin in biological fluids is closely related to human health. In this study, we developed a highly selective and robust assay to determine human serum albumin (HSA) in human plasma by combining chymotrypsin/trypsin digestion coupled with targeted LC-MS/MS technique. Human plasma samples were denatured, reduced, alkylated, and digested with both chymotrypsin and trypsin to generate surrogate peptides. A unique chymotryptic peptide (NAETF) arising from human serum albumin was finally selected for targeted LC-MS/MS detection and quantification. Numerous parameters related to the targeted LC-MS/MS assay were evaluated, including lower limit of quantitation (LLOQ), linearity range, enzyme digestion efficiency, accuracy and precision. The LC-MS/MS assay was linear in the concentration range 0.05-1 mg/mL with intra-day and inter-day precision <10.2% and accuracy ranging from -3.94% to 4.89%. The assay was successfully applied to determine HSA in 148 human plasma samples.