Discrepancy between two invasive blood pressure measurements in patients receiving intra-aortic balloon pump therapy
Lijuan Lu1, Shiyi Zhang2, Yu Zhang2
1Nursing Department, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, No. 1095, Jiefang Avenue, Wuhan, 430000, Hubei Province, China. 819081876@qq.com.
Background:
Hemodynamic monitoring is imperative for patients with cardiogenic shock undergoing Intra-aortic Balloon Pump (IABP) therapy. Blood pressure monitoring encompasses non-invasive, invasive peripheral arterial pressure (IPAP), and invasive central aortic pressure (ICAP) methods. However, marked disparities exist between IPAP and ICAP. This study examined the discrepancies between IPAP and ICAP and their clinical significance.
Methods:
A retrospective analysis was conducted on cardiogenic shock patients who underwent IABP therapy and were admitted to the Coronary Care Unit (CCU) of a tertiary hospital in China from March 2017 to November 2022. The Bland-Altman plot illustrated the discrepancy between IPAP and ICAP. A clinically significant difference between ICAP and IPAP measurements was defined as ≥ 10 mmHg, which could necessitate alterations in blood pressure management according to current guidelines that recommend maintaining a mean arterial pressure (MAP) ≥ 70 mmHg.
Results:
In total, 162 patients were included in the final analysis. In patients without vasopressors, the difference between ICAP and IPAP was 5.73 mmHg (95% limits of agreement [LOA], -16.98 to 28.44), whereas, in patients with vasopressors, it was 4.36 mmHg (95% LOA, -17.31 to 26.03). ICAP measurements exceeded IPAP in patients undergoing IABP therapy. However, the difference was not statistically significant between the two groups. Multivariate logistic regression revealed that higher serum lactate levels (Odds ratio [OR], 1.14; 95% confidence interval [CI], 1.03-1.27; p = 0.013) and age ≥ 60 years (OR, 13.20; 95% CI, 1.50-115.51; p = 0.020) were associated with an increased likelihood of a clinically significant MAP discrepancy. Conversely, a history of coronary heart disease was associated with a decreased likelihood (OR, 0.34; 95% CI, 0.13-0.90; p = 0.031).
Conclusions:
Notable discrepancies between ICAP and IPAP measurements exist in cardiogenic shock patients undergoing IABP therapy. ICAP exceeds IPAP, and factors such as age ≥ 60 years, elevated lactic acid levels, and absence of coronary heart disease contribute to this discrepancy. Enhanced vigilance is warranted for these patients, and the consideration of peripheral invasive monitoring in conjunction with IABP therapy is advised.
Related Concept Videos
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...