Nonlinear Gradient Field Mapping Using a Spherical Grid Phantom for 3 and 7 Tesla MR Imaging Systems Equipped with
Ryoichi Kose1, Katsumi Kose1, Koji Fujimoto2
1MRI simulations Inc.
Summary
High-performance MRI gradient coils often sacrifice spatial linearity. This study found vendor-provided correction formulas for magnetic field nonlinearity were insufficient, requiring verification with geometric phantoms.
Area of Science:
- Medical Imaging
- Magnetic Resonance Imaging (MRI)
- Gradient Coil Technology
Background:
- High-performance gradient coils are crucial for advanced MRI applications.
- Spatial linearity of gradient fields is often compromised in high-performance systems.
- Accurate magnetic field linearity is essential for precise image reconstruction.
Purpose of the Study:
- To measure and quantify the spatial nonlinearity of gradient fields in MRI systems.
- To evaluate the efficacy of vendor-provided geometric distortion correction methods.
- To develop and validate a method for correcting gradient field nonlinearity.
Main Methods:
- Utilized 3D gradient echo sequences and a spherical phantom with a lattice structure for measurements.
- Approximated spatial gradient field variations using 3rd order polynomials.
- Calculated polynomial coefficients via the steepest descent method.
- Corrected geometric distortion in 3D MR images using derived polynomials and compared with vendor methods.
Main Results:
- Quantified significant spatial nonlinearity in the gradient fields of two MRI systems.
- Demonstrated that vendor-provided harmonic function-based corrections were insufficient.
- The developed polynomial-based correction method showed improved accuracy.
Conclusions:
- Vendor-supplied nonlinearity correction formulae for MRI gradient fields require verification.
- Geometric phantoms are essential tools for assessing and correcting gradient field nonlinearity.
- Accurate correction of gradient field nonlinearity is critical for reliable 3D MR imaging.
Related Concept Videos
Magnetic Resonance Imaging
5.2K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
5.2K
Gradient and Del Operator
2.6K
In mathematics and physics, the gradient and del operator are fundamental concepts used to describe the behavior of functions and fields in space. The gradient is a mathematical operator that gives both the magnitude and direction of the maximum spatial rate of change. Consider a person standing on a mountain. The slope of the mountain at any given point is not defined unless it is quantified in a particular direction. For this reason, a "directional derivative" is defined, which is a vector...