Achieving Isotropic Super-Resolution with a Non-Isocentric Acquisition Geometry in a Next-Generation Tomosynthesis
Raymond J Acciavatti1, Chloe J Choi1, Trevor L Vent1
1University of Pennsylvania, Department of Radiology, 3400 Spruce Street, Philadelphia PA 19104.
Abstract:
We have constructed a prototype next-generation tomosynthesis (NGT) system that supports a non-isocentric acquisition geometry for digital breast tomosynthesis (DBT). In this geometry, the detector gradually descends in the superior-to-inferior direction. The aim of this work is to demonstrate that this geometry offers isotropic super-resolution (SR), unlike clinical DBT systems which are characterized by anisotropies in SR. To this end, a theoretical model of a sinusoidal test object was developed with frequency exceeding the alias frequency of the detector. We simulated two geometries: (1) a conventional geometry with a stationary detector, and (2) a non-isocentric geometry. The input frequency was varied over the full 360° range of angles in the plane of the object. To investigate whether SR was achieved, we calculated the Fourier transform of the reconstruction. The amplitude of the tallest peak below the alias frequency was measured relative to the peak at the input frequency. This ratio (termed the r-factor) should approach zero to achieve high-quality SR. In the conventional geometry, the r-factor was minimized (approaching zero) if the orientation of the frequency was parallel with the source motion, yet exceeded unity (prohibiting SR) in the orientation perpendicular to the source motion. However, in the non-isocentric geometry, the r-factor was minimized (approaching zero) for all orientations of the frequency, meaning SR was achieved isotropically. In summary, isotropic SR in DBT can be achieved using the non-isocentric acquisition geometry supported by the NGT system.
Related Concept Videos
Super-resolution Fluorescence Microscopy
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...