Accuracy of respiratory gas variables, substrate, and energy use from 15 CPET systems during simulated and human
Bas Van Hooren1, Tjeu Souren2, Bart C Bongers1,3
1Department of Nutrition and Movement Sciences, NUTRIM School of Nutrition and Translational Research in Metabolism, Maastricht University, Maastricht, The Netherlands.
Purpose:
Various systems are available for cardiopulmonary exercise testing (CPET), but their accuracy remains largely unexplored. We evaluate the accuracy of 15 popular CPET systems to assess respiratory variables, substrate use, and energy expenditure during simulated exercise. Cross-comparisons were also performed during human cycling experiments (i.e., verification of simulation findings), and between-session reliability was assessed for a subset of systems.
Methods:
A metabolic simulator was used to simulate breath-by-breath gas exchange, and the values measured by each system (minute ventilation [V̇E], breathing frequency [BF], oxygen uptake [V̇O2 ], carbon dioxide production [V̇CO2 ], respiratory exchange ratio [RER], energy from carbs and fats, and total energy expenditure) were compared to the simulated values to assess the accuracy. The following manufacturers (system) were assessed: COSMED (Quark CPET, K5), Cortex (MetaLyzer 3B, MetaMax 3B), Vyaire (Vyntus CPX, Oxycon Pro), Maastricht Instruments (Omnical), MGC Diagnostics (Ergocard Clinical, Ergocard Pro, Ultima), Ganshorn/Schiller (PowerCube Ergo), Geratherm (Ergostik), VO2master (VO2masterPro), PNOĒ (PNOĒ), and Calibre Biometrics (Calibre).
Results:
Absolute percentage errors during the simulations ranged from 1.15%-44.3% for V̇E, 1.05-3.79% for BF, 1.10%-13.3% for V̇O2 , 1.07%-18.3% for V̇CO2 , 0.62%-14.8% for RER, 5.52%-99.0% for Kcal from carbs, 5.13%-133% for Kcal from fats, and 0.59%-12.1% for total energy expenditure. Between-session variation ranged from 0.86%-21.0% for V̇O2 and 1.14%-20.2% for V̇CO2 , respectively.
Conclusion:
The error of respiratory gas variables, substrate, and energy use differed substantially between systems, with only a few systems demonstrating a consistent acceptable error. We extensively discuss the implications of our findings for clinicians, researchers and other CPET users.
More Related Videos
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
08:44Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...
Factors Affecting Respiration
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...