Measurement of Extracellular Ion Fluxes Using the Ion-selective Self-referencing Microelectrode Technique
Published on: May 3, 2015
Fluoride Determination in Milk, Soy, and Water-Based Products Using Ion-Selective Electrode and Direct Measurement
Renée M Erney1, Charles K Black1
1Abbott Nutrition, 3300 Stelzer Rd, Columbus, OH 43219, United States.
Background:
The AOAC Stakeholder Panel on Infant Formula and Adult Nutritionals issued a call for methods for the determination of fluoride in infant formula. Standard Method Performance Requirement (SMPR®) 2014.016 was approved.
Objective:
A single-laboratory validation (SLV) of a fluoride ion-selective electrode (F-ISE) method was completed.
Methods:
Five powder samples were reconstituted, and hydrochloric acid was added to dissolve and release any bound fluoride. Samples included infant and adult nutritional products made from milk, whey, or soy, containing intact, partially hydrolyzed, or hydrolyzed proteins. Sodium citrate buffer was added to complex any interfering ions and adjust pH and ionic strength. Samples were analyzed in duplicate over 6 days using F-ISE.
Results:
Results were calculated based on one of two least-squares (log fit) standard curves (0.02 to 0.1 µg/100 g or 0.1 to 2.0 µg/100 g). The LOQ for the method was determined to be 3.3 µg/100 g, which was far below the minimum analytical value of 30 µg/100 g required in the SMPR. The fluoride concentrations ranged from approximately 3 to 14 µg/100 g reconstituted powder and the intermediate precision, RSDr, ranged from 0.7 to 12.4%. Although all of these results are out of scope for the SMPR due to their low fluoride, it should be noted that the RSDr for formulas with fluoride concentrations ranging from approximately 9 to 14 µg/100 g ranged from 0.7 to 4.3%, and the RSDr for the two formulas with fluoride concentrations at or below the theoretical detection limit of the method were only approximately10 and 12.4%. Recoveries ranged from 94 to 98% for samples spiked at three levels with NIST-traceable standard solutions, meeting the SMPR.
Conclusion:
The F-ISE method meets the SMPR and was adopted as a First Action Official MethodSM.
Highlights:
The fluoride ion-selective electrode method presented here is affordable and easy to run. The addition of acid to the samples dissolves any minerals which may bind the fluoride.
Related Concept Videos
EDTA: Indirect and Alkalimetric Titration
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Complexometric Titration: Ligands
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides...
Ion-Exchange Chromatography