Use of Two Intracorporeal Ventricular Assist Devices As a Total Artificial Heart
Published on: May 11, 2018
Balancing the ventricular outputs of pulsatile total artificial hearts
Oskar J Gülcher1,2, Annemijn Vis1, Mathias Peirlinck2
1Department of Cardiothoracic Surgery, Amsterdam University Medical Centers, Location University of Amsterdam, Amsterdam, The Netherlands.
Background:
Maintaining balanced left and right cardiac outputs in a total artificial heart (TAH) is challenging due to the need for continuous adaptation to changing hemodynamic conditions. Proper balance in ventricular outputs of the left and right ventricles requires a preload-sensitive response and mechanisms to address the higher volumetric efficiency of the right ventricle.
Methods:
This review provides a comprehensive overview of various methods used to balance left and right ventricular outputs in pulsatile total artificial hearts, categorized based on their actuation mechanism.
Results:
Reported strategies include incorporating compliant materials and/or air cushions inside the ventricles, employing active control mechanisms to regulate ventricular filling state, and utilizing various shunts (such as hydraulic or intra-atrial shunts). Furthermore, reducing right ventricular stroke volume compared to the left often serves to balance the ventricular outputs. Individually controlled actuation of both ventricles in a pulsatile TAH seems to be the simplest and most effective way to achieve proper preload sensitivity and left-right output balance. Pneumatically actuated TAHs have the advantage to respond passively to preload changes.
Conclusion:
Therefore, a pneumatic TAH that comprises two individually actuated ventricles appears to be a more desirable option-both in terms of simplicity and efficacy-to respond to changing hemodynamic conditions.
Related Concept Videos
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Heart Failure VI: Adjunct Therapies
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...