Automated Slide Scanning and Segmentation in Fluorescently-labeled Tissues Using a Widefield High-content Analysis System
Published on: May 3, 2018
A dynamic parallel image acquisition method for slide scanning process
Bowen Zhong1, Zhan Liao1, Xiaopeng Hao1
1The College of Mechanical and Electrical Engineering, Soochow University, Suzhou 215000, China.
Abstract:
Automated microscope systems have played an important role in the screening of numerous diseases. However, it is a very time-consuming process to continuously acquire images under the high magnification objective lens. This paper proposes a dynamic parallel image acquisition method, which can greatly improve image acquisition speed. Due to the relative motion between the x-y stage and the camera, some of the captured images have motion blur To this end, we also designed a motor variable speed motion curve to ensure the quality of the collected images. The experimental results show that the traditional image scanning mode needs 47.3 ms to obtain continuous microscopic images, while the dynamic parallel image acquisition method only needs 25.4 ms, which improves the acquisition speed without affecting the clarity of the acquired images.