Diffuse photon remission associated with the center-illuminated-area-detection geometry. II. Approach to the
Abstract:
This part proposes a model of time-dependent diffuse photon remission for the center-illuminated-area-detection (CIAD) geometry, by virtue of area integration of the radially resolved time-dependent diffuse photon remission formulated with the master-slave dual-source scheme demonstrated in Part I for steady-state measurements. The time-domain model is assessed against Monte Carlo (MC) simulations limiting to only the Heyney-Greenstein scattering phase function for CIAD of physical scales and medium properties relevant to single-fiber reflectance (SfR) and over a 2 ns duration, in compliance with the timespan of the only experimental report of SfR demonstrated with a 50 µm gradient index fiber. The time-domain model-MC assessments are carried out for an absorption coefficient ranging three orders of magnitude over [0.001,0.01,0.1,1]m m -1 at a fixed scattering, and a reduced scattering coefficient ranging three orders of magnitude over [0.01,0.1,1,10]m m -1 at a fixed absorption, among others. Photons of shorter and longer propagation times, relative to the diameter of the area of collection, respond differently to the scattering and absorption changes. Limited comparisons of MC between CIAD and a top-hat geometry as the idealization of SfR reveal that the time-domain photon remissions of the two geometries differ appreciably in only the early arriving photons.