Subcutaneous Implantable Cardioverter-Defibrillator Lead Dislodgement Masquerading as Left Shoulder Pain
Yashitha Chirumamilla1, Deepesh Yadav2,1, Ghassan Bachuwa1
1Internal Medicine, Hurley Medical Center, Flint, USA.
Abstract:
Implantable cardioverter-defibrillators (ICDs) have demonstrated efficacy in the prevention of sudden cardiac death secondary to cardiac arrhythmias in eligible patients. Complications with the subcutaneous ICD (S-ICD) are rarer than with the transvenous ICD but do still exist. Our patient presented four weeks after the insertion of S-ICD with complaints of left shoulder pain radiating to the chest wall and swelling over the S-ICD site. He was initially treated for rotator cuff injury and subacromial impingement syndrome but upon obtaining chest radiography was found to have a lead displacement traversing the splenic flexure of the colon. The patient was managed by a treatment team involving cardiology, surgery, and infectious disease and underwent S-ICD removal, exploratory laparotomy with splenic flexure mobilization, and completion of a four-week antibiotic course ultimately leading to reimplantation of S-ICD.
Related Concept Videos
Pericarditis II: Clinical Features and Diagnostic Tests
Pericarditis IV: Nursing Management
Angina III: Clinical Manifestations and Assessment
Cardiac Catheterization III: Left Heart Catheterization
Cardiomyopathy V: Interprofessional Care
Cardiomyopathy II: Dilated Cardiomyopathy