Related Experiment Video
Updated: Jul 16, 2025

3D Planning and Printing of Patient Specific Implants for Reconstruction of Bony Defects
Published on: August 4, 2020
Stability of Three-Dimensional Printed Custom-Made Metaphyseal Cone for Tibial Bone Defects Reconstruction: A Finite
Xinguang Wang1,2, Xinyu Li3, Cheng Wang1,2
1Department of Orthopaedics, Peking University Third Hospital, Beijing, China.
Objectives:
The reconstruction of bone defects in tibial revision knee arthroplasty is challenging. In this study, we evaluated the primary stability of a novel three-dimensional (3D)-printed custom-made metaphyseal cone for Anderson Orthopedic Research Institute (AORI) IIb or III bone defect reconstruction in tibial revision knee arthroplasty using the combination of finite-element analysis and biomechanical experiments.
Methods:
In the finite-element analysis, AORI II b and III medial tibial bone defects were designed at varying depths. A novel 3D-printed custom-made metaphyseal cone was designed and used to reconstruct the bone defect with or without a stem in simulated revision total knee arthroplasty (RTKA). A no-stem group and a stem group were established (based on whether a stem was used or not). Von Mises stress and micromotion were calculated with varying depths of bone defects, ranging from 5 mm to 35 mm, and then micromotions at the bone-implant interface were calculated and compared with the critical value of 150 μm. In the biomechanical experiment, the no-stem group was used, and the same bone defects were made in four synthetic tibias using patient-specific instruments. Micromotions at the bone-implant interface were investigated using a non-contact optical digital image correlation system and compared with the critical value of 150 μm.
Results:
When the bone defect was <30 mm, micromotions at the bone-implant interface in the finite-element analysis were all below 150 μm both in the stem groups and no-stem groups, whereas those in the biomechanical experiment were also below 150 μm in the no-stem group.
Conclusions:
The 3D-printed custom-made metaphyseal cone in RTKA has excellent primary stability and does not require stems in reconstructing tibial AORI type IIb or III bone defects with a depth of <30 mm.