Related Experiment Video
Updated: Jul 16, 2025

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Fundamental research for dose control during supine dental panoramic radiography
Atsuharu Nitanda1, Atsushi Iwawaki1, Yusei Otaka2
1Division of Forensic Odontology, Department of Diagnostic and Therapeutic Sciences, Meikai University School of Dentistry, Saitama, Japan.
Objective:
This study aimed to control radiation doses when using a portable supine dental panoramic radiography system by measured the scattered doses.
Method:
The study used LPX7007 (Asahi Roentgen) for the panoramic radiography system. The subjects comprised a cylinder phantom (QualitA) and a RANDO Phantom (Alderson). The semiconductor dosimeter was an X2 survey sensor (RaySafe). The phantom was set at a height of 1 m from the floor, and the sensor was set at 1 m from the floor at the genital level and 1.5 m at the lens level. Measurements were taken at 30°intervals clockwise from 0°at distances of 0.5 m and 1 m in radius around the phantom. The occupational exposure range was defined as 0 ± 30° and the public exposure range was defined as the occupational exposure range and 30° to 150° and 210° to 330° as the public exposure range.
Result:
The highest doses were observed in the 120° and 240° directions, and the lowest in 0° ± 30° range. The lowest limit number of images taken in the occupational exposure range was 130 images at a distance of 0.5 m, 452 images at 1 m at the lens level for the cylinder phantom, and 320 images at 0.5 m and 1098 images at 1 m for the RANDO Phantom. In the public exposure range at the genital level, there was one image at 0.5 m and six images at 1 m for the cylinder phantom, and two images at 0.5 m and eight images at 1 m for the RANDO Phantom.
Conclusion:
We found that radiation exposure can be reduced by keeping a distance from the subject, avoiding working at 120° and 240° and staying within 0° ± 30° behind the panoramic radiography system.
More Related Videos
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
X-ray Imaging