Related Experiment Video
Updated: Jul 16, 2025

Experimental Methods for Trapping Ions Using Microfabricated Surface Ion Traps
Published on: August 17, 2017
A versatile 16-pole ion trap setup for investigating photophysics of biomolecular ions
M Salvi1, N N Uma1, Hemanth Dinesan2
1Department of Physics and the Center for Atomic, Molecular, and Optical Sciences & Technologies (CAMOST), Indian Institute of Science Education and Research Tirupati, AP, Tirupati 517507, India.
Abstract:
A linear 16-pole ion trap-based experimental setup has been designed, implemented, and characterized to investigate the photophysics of biomolecules in the gas phase. Electrospray ionization is employed to generate the ions in the gas phase at atmospheric pressure. The voltage configuration on the ion funnel, the ion optic device in the first vacuum interface, is used to control the energy of the ions. A home-built quadrupole mass-filter is utilized for the mass-selection of the ions of interest. A 16-pole ion trap designed and built in-house is implemented for ion trapping. The instrument's versatility and capability are showcased by demonstrating the fragmentation patterns of protonated and deprotonated tryptophan, as well as describing the photodetachment decay of deprotonated indole.