Validation of an integrated pressure level measured earmold wideband real-ear-to-coupler difference measurement
Matthew Urichuk1,2, David Purcell1,2,3, Prudence Allen1,2,3
1Faculty of Health Sciences, School of Communication Sciences and Disorders, Western University, London, Ontario, Canada.
Objective:
To validate measurement of predicted earmold wideband real-ear-to-coupler difference (wRECD) using an integrated pressure level (IPL) calibrated transducer and the incorporation of an acoustically measured tubing length correction.
Design:
Unilateral earmold SPL wRECD using varied hearing aid tubing length and the proposed predicted earmold IPL wRECD measurement procedure were completed on all participants and compared.
Study Sample:
22 normal hearing adults with normal middle ear status were recruited.
Results:
There were no clinically significant differences between probe-microphone and predicted earmold IPL wRECD measurements between 500 and 2500 Hz. Above 5000 Hz, the predicted earmold IPL wRECD exceeded earmold SPL wRECDs due to lack of standing wave interference. Test-retest reliability of IPL wRECD measurement exceeded the reliability of earmold SPL wRECD measurement across all assessed frequencies, with the greatest improvements in the high frequencies. The acoustically measured tubing length correction largely accounted for acoustic effects of the participant's earmold.
Conclusions:
IPL-based measurements provide a promising alternative to probe-microphone earmold wRECD procedures. Predicted earmold IPL wRECD is measured without probe-microphone placement, agrees well with earmold SPL wRECDs and is expected to extend the valid bandwidth of wRECD measurement.
Related Concept Videos
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Measurement of Blood Pressure
Pressure Gauges
Intensity and Pressure of Sound Waves
Unlike the time average of a sinusoidal term, which is zero since it is positive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...