Determination of sources of error and improvement in accuracy of left ventricular mass measurement by
Najmul Siddiqi1, Sunil Vasireddi2, Prasongchai Sattayaprasert2
1Heart and Vascular Center, MetroHealth Campus of Case Western Reserve University, 2500 MetroHealth Drive, Cleveland, OH, 44109, USA. siddiqi.najmul@gmail.com.
Abstract:
Abnormal Left ventricular mass (LVM) prognosticates adverse cardiovascular events. Conventionally, LVM measured by echo assumes a prolate ellipsoid (PE) shape; however, it poorly correlates with reference standard of cardiac magnetic resonance imaging (CMR) derived LVM. PE model assumes LVL = 2 × LVID. We developed a new echo LVM formula based on LV length and tested for accuracy against CMR. A retrospective study of consecutive patients with an echocardiogram and CMR within 3 months. Derivation (n = 170) and validation cohorts (n = 54) were used to test the new formula. Following analysis of correlation of interventricular septum (IVS), LV internal dimension (LVID), posterior wall (PW) and LVL between echo and CMR, a novel paraboloid-shape linear regression (PLR) model was derived. LVM by both models were compared to CMR. Poor correlation observed between actual and assumed LVL (0.52 with CMR; 0.44 with echo). Strong correlation was noted between echo and CMR measured LVL, LVID, IVS (r > 0.80) and a moderate correlation with PW (r = 0.62). Strong correlation of LVL was harnessed to develop PLR model, which significantly decreased paired error in derivation cohort (from 64 ± 42 to 22 ± 21 gm) and validation cohort (from 63 ± 46 to 25 ± 18 gm). Furthermore, it demonstrates significant reduction in absolute, relative errors and variability along with superior correlation in both cohorts. Between echo and CMR, LVL demonstrates one of the best correlation among LV dimensions. The assumption, LVL = 2 × LVID appears inaccurate. PLR model incorporates LVL and significantly improves accuracy, reduces variability of LVM.
More Related Videos
09:05Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Errors occurring during blood pressure monitoring
Several factors...
Random and Systematic Errors
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...