Use of Digital Stethoscope to Measure Heart Rate in Birds: Comparison of Different Counting Methods Using
Ciro Cococcetta1, Thomas Coutant2, Sophie Bagur3
1Service des Nouveaux Animaux de Compagnie Centre Hospitalier Vétérinaire Frégis, 94250 Gentilly, France, cirocococ@ gmail.com.
Abstract:
The high cardiac contractility of birds poses a challenge to traditional cardiac auscultation, particularly for the accurate determination of heart rate (HR). The objectives of this study were to 1) evaluate the feasibility of using phonocardiograms of adequate length and quality to assess HR in different avian species with a commercially available digital stethoscope, 2) compare 5 counting methods, including 2 direct reading methods (manual counting and using a semiautomatic computerized algorithm as a reference method) and 3 listening methods (progressive mental counting, counting by 10s, and counting with a smartphone application tap counter), and 3) obtain the HR in selected birds and identify a correlation between body weight and HR in different avian species. An inverse correlation on a logarithmic scale was identified between the mean body weight and HR in 60 different bird species (n = 211; R = -0.72, P < 0.0001). Manual reading of phonocardiograms was the most reliable method and had the highest agreement with the reference method; this was followed by the counting by 10s method, the tapping method, and the progressive mental counting method, which was the least reliable. The agreement levels for the different methods were comparable for HRs <200 beats per minute (bpm) in birds weighing >1 kg. For HRs >500 bpm in birds weighing <150 g, only the reading method maintained a good agreement level. A digital stethoscope can be a useful tool for accurately determining the HR in birds, including very small species with high HRs.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V)...
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.