Carbon monoxide analysis: a comparison of two co-oximeters and headspace gas chromatography
Abstract:
Three methods (IL-182 Co-Oximeter, IL-282 Co-Oximeter, and headspace gas chromatography) for the analysis of carbon monoxide in postmortem blood were studied and compared using a prepared reference standard, Quantra control materials, and 62 postmortem blood specimens. The methods compared favorably with one another. The linear regression equations for the 62 postmortem blood samples (range = 1.0 to 86% saturation) were: IL-282 vs. IL-182, y = 1.11x - 3.10, r = 0.981; IL-182 vs. GC, y = 0.88x + 2.97, r = 0.973; IL-282 vs. GC, y = 1.00x - 1.24, r = 0.986.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II