Commissural alignment in the Evolut TAVR procedure: conventional versus hat marker-guided shaft rotation methods
Yutaka Konami1,2, Tomohiro Sakamoto1, Hiroto Suzuyama1
1Division of Cardiology, Cardiovascular Center, Saiseikai Kumamoto Hospital, Kumamoto, Japan.
Background:
Coronary cannulation after TAVR is sometimes difficult due to an overlap between native and neo-commissures, especially in Evolut devices with a supra-annular position. The Evolut C-tab corresponds to a neo-commissure, and the hat marker is in a fixed position. Therefore, the orientation of the hat marker can be adjusted to minimise overlaps.
Aims:
We investigated whether the HAt marker-guided SHaft rotation method (HASH, stylised as the #rotation method) is effective in facilitating coronary artery access after transcatheter aortic valve replacement (TAVR) with an Evolut system.
Methods:
We retrospectively analysed 95 patients who underwent electrocardiogram-gated cardiac computed tomography after TAVR. In the #rotation method, the hat marker of the delivery catheter was adjusted to face the greater curvature of the descending thoracic aorta in the left anterior oblique view. Its orientation was maintained while the system passed through the aortic arch.
Results:
In total, 60 and 35 patients underwent TAVR with the #rotation and non-#rotation methods, respectively. A ±15° angle between the native and neo-commissures was more frequent in the #rotation group (p=0.001). Favourable angles and appropriate frame orientation for access to the left coronary artery were significantly more frequent in the #rotation group than in the non-#rotation group (p<0.001 and p=0.001). Although the #rotation method showed a higher rate of favourable angles and frames in the right coronary artery, statistically significant differences were not found.
Conclusions:
The #rotation method is useful for improving commissural post alignment in TAVR with Evolut devices, especially in the ostium of the left coronary artery.