Related Experiment Video
Updated: Jul 16, 2025

Novel and Innovative Hybrid Technique for Type A Aortic Dissection
Published on: March 28, 2025
A Case of Cardiac Arrest Due to Transcatheter Aortic Valve Infolding
Fawaz Mohammed1, James C Gubitosa2, Travis R Huffman2
1Internal Medicine, University of Kentucky College of Medicine, Bowling Green, USA.
Abstract:
Prosthetic valvular infolding during transcatheter aortic valve implantation (TAVI) is an under-recognized yet significant complication that can occur. Here, we describe the case of a 61-year-old male with a history of heart failure with reduced ejection fraction (HFrEF) and low-flow, low-gradient severe aortic valve stenosis of a bicuspid aortic valve who presented to undergo TAVI. During the procedure, repositioning of the valve resulted in prosthetic valvular infolding and resultant severe aortic regurgitation (AR), culminating in cardiac arrest. Swift balloon valvuloplasty corrected the valve geometry and eliminated any AR, allowing hemodynamic recovery and completion of the procedure. Our case and review highlight methods, both angiographic and echocardiographic, to recognize prosthetic valvular infolding the moment it presents, as well as strategies to correct the infolding with minimal detriment to the patient.