Determinants of apnea-hypopnea index variability during home sleep testing
Jean-Benoît Martinot1, Nhat-Nam Le-Dong2, Renaud Tamisier3
1Sleep Laboratory, CHU Université Catholique de Louvain (UCL), Namur Site Sainte-Elisabeth, Namur, Belgium; Institute of Experimental and Clinical Research, UCL Bruxelles Woluwe, Brussels, Belgium.
Background:
A single-night attended in-laboratory polysomnography or home sleep testing are common approaches for obstructive sleep apnea (OSA) diagnosis. However, internight variability in apnea-hypopnea index value is common, and may result in misclassification of OSA severity and inapropriate treatment decisions.
Objective:
To investigate factors determining short-term apnea-hypopnea index variability using multi-night automated home sleep testing, and to determine how this variability impacts clinical decisions.
Patients/Methods:
Adults with suspected OSA who successfully performed three home sleep tests using measurements of mandibular jaw movements (Sunrise, Namur, Belgium) combined with automated machine learning analysis were enrolled. Data analysis included principal component analysis, generalized estimating equation regression and qualitative agreement analysis.
Results:
160 individuals who performed three sleep tests over a mean of 8.78 ± 8.48 days were included. The apnea-hypopnea index varied by -0.88 events/h (5th-95th percentile range: -14.33 to 9.72 events/h). Based on a single-night recording, rates of overtreatment and undertreatment would have been of 13.5% and 6.0%, respectively. Regression analysis adjusted for age, sex, body mass index, total sleep time, and time between home sleep tests showed that time spent in deep non-rapid eye movement sleep and with head in supine position were independent significant predictors of the apnea-hypopnea index variability.
Conclusions:
At the individual level, short-term internight variability in the apnea-hypopnea index was significantly associated with time spent in deep non-rapid eye movement sleep and head in supine position. Clinical decisions based on a single-night testing may lead to errors in OSA severity classification and incorrect therapeutic decisions.
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Respiratory Volumes and Capacities I
Other Factors Affecting Respiration Centers
However, the ability to hold one's breath voluntarily is not limitless. When the CO2 concentration in the blood reaches a critical...
Factors Affecting Respiration
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...