Extracorporeal Membrane Oxygenation Physiological Factors Influence Pulse Oximetry and Arterial Oxygen Saturation
Andrew Kalra1, Benjamin L Shou2, David Zhao3
1Division of Cardiac Surgery, Department of Surgery, The Johns Hopkins Hospital, Baltimore, Maryland; Sidney Kimmel Medical College, Thomas Jefferson University, Philadelphia, Pennsylvania.
Background:
Cannulation strategy, vasopressors, and hemolysis are important physiological factors that influence hemodynamics in extracorporeal membrane oxygenation (ECMO). We hypothesized these factors influence the discrepancy between oxygen saturation measured by pulse oximetry (Spo2) and arterial blood gas (Sao2) in patients on ECMO.
Methods:
We retrospectively analyzed adults (aged ≥18 years) on venoarterial or venovenous ECMO at a tertiary academic ECMO center. Spo2-Sao2 pairs with oxygen saturation ≥70% and measured within 10 minutes were included. Occult hypoxemia was defined as Sao2 ≤88% with a time-matched Spo2 ≥92%. Adjusted linear mixed-effects modeling was used to assess the Spo2-Sao2 discrepancy with preselected demographics and time-matched laboratory variables. Vasopressor use was quantified by vasopressor dose equivalences.
Results:
Of 139 venoarterial-ECMO and 88 venovenous-ECMO patients, we examined 20,053 Spo2-Sao2 pairs. The Spo2-Sao2 discrepancy was greater in venovenous-ECMO (1.15%) vs venoarterial-ECMO (-0.35%, P < .001). Overall, 81 patients (35%) experienced occult hypoxemia during ECMO. Occult hypoxemia was more common in venovenous-ECMO (65%) than in venoarterial-ECMO (17%, P < .001). In linear mixed-effects modeling, Spo2 underestimated Sao2 by 9.48% in central vs peripheral venoarterial-ECMO (95% CI, -17.1% to -1.79%; P = .02). Higher vasopressor dose equivalences significantly worsened the Spo2-Sao2 discrepancy (P < .001). In linear mixed-effects modeling, Spo2 overestimated Sao2 by 25.43% in single lumen-cannulated vs double lumen-cannulated venovenous-ECMO (95% CI, 5.27%-45.6%; P = .03). Higher vasopressor dose equivalences and lactate dehydrogenase levels significantly worsened the Spo2-Sao2 discrepancy (P < .001).
Conclusions:
Venovenous-ECMO patients are at higher risk for occult hypoxemia compared with venoarterial-ECMO. A higher vasopressor requirement and different cannulation strategies (central venoarterial-ECMO; single-lumen venovenous-ECMO) were significant factors for clinically significant Spo2-Sao2 discrepancy in both ECMO modes.
More Related Videos
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Respiration and Gaseous Exchange
Respiration involves the exchange of gases, especially oxygen (O2) and carbon dioxide (CO2), between the alveoli and body cells, a process facilitated by blood circulation. As a result, the cardiovascular system, which involves...
Factors Affecting Respiration