Related Experiment Video
Updated: Jul 15, 2025

Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
Assessment of High-Variability Speech Recognition in Adult Cochlear Implant Users using PRESTO
Terrin N Tamati1, Kathleen F Faulkner2, David B Pisoni3
1Department of Otolaryngology, Vanderbilt University Medical Center, Nashville, Tennessee.
Background:
Speech recognition in adult cochlear implant (CI) users is typically assessed using sentence materials with low talker variability. Little is known about the effects of talker variability on speech recognition in adult CI users, the factors underlying individual differences in speech recognition with high talker variability, or how sentence materials with high talker variability could be utilized clinically.
Purpose:
To examine the effects of talker variability on sentence recognition in adult CI users, using sentences from the Perceptually Robust English Sentence Test Open-Set (PRESTO), and to examine the relation between working memory capacity and high-variability speech recognition.
Research Design:
Postlingually deafened adult CI users and adults with self-reported normal hearing (NH) under CI simulation completed sentence recognition tests that contained varying levels of talker variability, including Hearing in Noise Test (HINT; low-variability), AzBio (moderate-variability), and PRESTO sentences (high-variability). The tasks were completed in both quiet and multitalker babble (MTB). For the adult CI users only, the relation between sentence recognition accuracy and working memory capacity was assessed.
Study Sample:
Twenty postlingually deafened adult CI users and 35 NH adults under 8-channel acoustic noise-vocoder simulations of CI hearing.
Results:
In both CI and NH groups, performance decreased as a function of increased talker variability, with the best scores obtained on HINT (low-variability), then AzBio (moderate-variability), followed by PRESTO (high-variability) in quiet. In MTB, performance was significantly lower on PRESTO sentences, compared with HINT and AzBio sentences, which were not significantly different. Working memory capacity in the CI users was related to sentence recognition accuracy across all materials and conditions.
Conclusion:
Findings from the current study suggest that the increased talker variability in the PRESTO sentence materials has a detrimental effect on speech recognition in both adult CI users and NH listeners under CI simulation, particularly when speech is further degraded by MTB. For adult CI users, working memory capacity contributes to speech recognition abilities. Sentence recognition testing with high-variability, multitalker materials, as in PRESTO, provides robust assessment of speech recognition abilities for research and clinical application, generating a wide range of scores for evaluating individual differences without ceiling effects when compared with conventional low-variability sentences.
More Related Videos
05:48Author Spotlight: Investigating the Impact of Emotional Prosodies on Voice Recognition and Perception
Published on: August 9, 2024
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Related Concept Videos
Auditory Perception
Perceiving Loudness, Pitch, and Location
Place theory, or place coding, suggests that different pitches are heard because various sound waves activate specific locations along the cochlea's basilar membrane. The brain determines the pitch of a sound by...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...