Related Experiment Video
Updated: Jul 15, 2025

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Bimodular femoral stems in primary total hip arthroplasty
Jan Zajc1,2, Samo K Fokter1,2
1Clinical Department of Orthopedic Surgery, University Medical Center Maribor, Maribor, Slovenia.
Introduction:
This review critically examines the efficacy of dual-modular stems in primary total hip arthroplasty. Given the variability and non-comparability of certain femoral stem designs and stem-neck couplings, with some even being withdrawn from the market, this review offers an in-depth analysis of predominant implant performances.
Areas Covered:
The paper explores a brief historical summary related to dual-modular stems, including the complications associated with their use, diagnostic tools for evaluation, analysis of both recalled and currently available models, as well as alternative therapeutic options. This information is pertinent for both clinical and research domains.
Expert Opinion:
While dual-modular systems were initially touted to offer several advantages, the evidence substantiating these benefits has been ambiguous. Further, these systems introduce the risk of alternative complications. In specific cases involving patients with developmental hip dysplasia and certain proximal femoral deformities requiring complex reconstructions, dual-modular systems might be relevant. Nonetheless, the use of long interchangeable necks in patients with a body mass index above 30 kg/m2 is discouraged, and pairing a long varus-oriented neck with an extra-long femoral head should be avoided in all patients.
More Related Videos
07:45The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
09:51The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022