Related Experiment Video
Updated: Jul 15, 2025

CMAP Scan MUNE MScan - A Novel Motor Unit Number Estimation MUNE Method
Published on: June 7, 2018
Sensitivity Analysis of CMAP Scan Step Index to Different Stimulation Parameters and Examination of Muscles Affected
Abstract:
Step index (STEPIX) is a recently developed compound muscle action potential (CMAP) scan method for evaluating motor unit loss and remodeling changes. This study investigates the influence of different stimulation parameters during CMAP scan on STEPIX and its examination of muscles affected by spinal cord injury (SCI). CMAP scan of the first dorsal interosseous (FDI) muscle was performed using different stimulus pulse widths (0.1 ms, 0.2 ms) and different numbers of stimuli (500, 1000) in 12 neurologically intact subjects. STEPIX was derived from each CMAP scan of all subjects. A significantly higher STEPIX was obtained using 1000 stimuli than 500 stimuli, while no significant difference in STEPIX was observed using 0.1 and 0.2 ms stimulus pulse widths. STEPIX was further applied to process CMAP scans of the FDI muscle from 13 tetraplegia and 13 healthy control subjects using the same stimulation parameter setting (0.1 ms, 500 stimuli), along with other methods including MScanFit motor unit number estimation (MUNE) and D50. STEPIX was significantly lower for the SCI subjects compared with the healthy control subjects. STEPIX was significantly correlated with MscanFit MUNE and D50, but had a smaller relative width of the overlapping zone (WOZ%) between tetraplegic and healthy control groups compared with MScanFit MUNE and D50. The findings of the study highlight the importance of maintaining a consistent stimulation parameter setting in CMAP scan studies and confirm the usefulness of STEPIX as a convenient CMAP scan parameter for examination of motor unit number changes.
More Related Videos
09:07Electrophysiological Motor Unit Number Estimation MUNE Measuring Compound Muscle Action Potential CMAP in Mouse Hindlimb Muscles
Published on: September 25, 2015
07:33Measuring Contralateral Silent Period Induced by Single-Pulse Transcranial Magnetic Stimulation to Investigate M1 Corticospinal Inhibition
Published on: August 23, 2022