Related Experiment Video
Updated: Jul 15, 2025

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Quantification of Drospirenone- and Ethinyl Estradiol-Related Impurities in a Combined Pharmaceutical Dosage Form by
Srinivasa Reddy Chinta1,2, Vaishnavi Chintala3, Vishnu Nandimalla2
1Novast Laboratories Ltd, Department of Quality Operations, Guangxing Rd, Free Trade Zone, NETDA, Nantong 226009, China.
Background:
The estimation of drugs containing drospirenone (DRSP) and ethinyl estradiol (EE), and their related impurities, in low-dose oral contraceptive drug products is an extremely challenging target. The proposed research sought to develop and validate a stability-indicating method for quantifying drug substances and their related impurities in tablet formulation.
Objective:
To develop and validate a simple, specific, accurate, precise, and stability-indicating reverse-phase (RP)-HPLC method for quantification of DRSP, EE, and their impurities in accordance with International Conference on Harmonisation (ICH) guidelines.
Method:
The separation was achieved using an Agilent Zorbax SB C18 column (4.6 mm × 250 mm, 5 µm) with a detection wavelength of 215 nm and mobile phases A (100% acetonitrile) and B (acetonitrile-water, 1 + 3, v/v) at a flow rate of 1.3 mL/min and a column temperature of 40°C.
Results:
The recovery study of each impurity was conducted in the range of 24 to 72 µg/mL for DRSP-related impurities and 0.2 to 0.6 µg/mL for EE-related impurities with respect to the specification limit. A linearity study was conducted over a range of 1.5 to 90 µg/mL for DRSP and DRSP-related impurities, and 0.125 to 0.75 µg/mL for EE-related impurities. A Quality by Design (QbD) study demonstrated the method's robustness.
Conclusions:
As per current guidelines, a stability-indicating method has been developed for the determination of impurities in DRSP/EE film-coated tablets. A QbD-based robustness test was performed and the method was found to be robust.
Highlights:
An accurate, precise, stability-indicating, gradient RP-HPLC method has been developed and validated to determine DRSP, EE, and nine related impurities in tablet formulation. A QbD technique was used to establish a robustness study.
More Related Videos
07:00Failure of Cleaning Verification in Pharmaceutical Industry Due to Uncleanliness of Stainless Steel Surface
Published on: August 11, 2017
07:29HPLC Coupled with Chemical Fingerprinting for Multi-Pattern Recognition for Identifying the Authenticity of Clematidis Armandii Caulis
Published on: November 11, 2022