Water interference in the chromatographic analysis of exhaled breath samples: Challenges and mitigation strategies
Konstantin Tartakovsky1, Shahar Geller1, Shmuel Rozenfeld1
1Scent Medical Technologies, Rehovot, 7670107, Israel.
Abstract:
This study demonstrates the adverse effects of water in exhaled breath samples on the accuracy of breath biomarker analysis when using gas chromatography. The presence of water in exhaled breath significantly modifies the retention times and peak areas of compounds, particularly for low-boiling, early eluting compounds. To tackle this issue, a two-step approach is introduced. The process begins with thorough desorption of the sorbent tube using a high split ratio and a short analysis duration, followed by a secondary analysis of the same tube. The efficacy of the new, straightforward approach was illustrated using humid breath samples and 57 compound standard mixture. This study highlights the importance of proper sample pretreatment and analysis to ensure reliable and accurate results in clinical research.
Related Concept Videos
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences