Related Experiment Video
Updated: Jul 15, 2025

Three-Dimensional Preoperative Virtual Planning in Derotational Proximal Femoral Osteotomy
Published on: February 17, 2023
The use of an anatomic tibial baseplate in total knee arthroplasty optimizes coverage without compromising rotation:
Jade Pei Yuik Ho1, Seong Yun Park2, Hee Seung Nam2
1Department of Orthopaedic Surgery, Seoul National University College of Medicine, Seoul National University Bundang Hospital, South Korea; Department of Orthopaedic Surgery, Kuala Lumpur General Hospital, Malaysia.
Background:
The aims of this study were (1) to compare in vivo coverage and rotational alignment of 2 tibial component designs: anatomic and symmetrical; and (2) to determine if coronal deformity and tibial torsion were related to rotation and coverage.
Methods:
Postoperative CT scans of 200 propensity score-matched patients who underwent TKA with either an anatomic (ATC) or symmetrical tibia component (STC) were analyzed. Rotation was measured using four axes: surgical transepicondylar axis (sTEA), Berger's protocol, medio-lateral (ML) axis and posterior borders of the tibial plateau, while coverage was assessed by measuring fit and surface area. The relationship between coronal deformity, tibial torsion, rotation, and coverage was investigated.
Results:
Overall, STCs had more internal rotation when measured using the sTEA (-0.6° ± 3.5 vs 0.5° ± 3.6, p = 0.03), Berger's protocol (-21.6° ± 7.1 vs -17.9° ± 6.2, p = 0.000) and ML axes (2.9° ± 3.9 vs 8.1° ± 5.1, p = 0.000) compared to ATCs. STCs also had more posteromedial underhang (-3.3 mm ± 2.4 mm vs -1.7 mm ± 2.5 mm, p = 0.000) but smaller change in tibial torsion postoperatively (-18.4° ± 9.9° vs -13.1° ± 9.4°, p = 0.000). Tibial torsion was more pronounced in valgus than varus knees both preoperative (-25.4° ± 6.5° vs -20.2° ± 9.3°, p = 0.02) and postoperatively (-19.7° ± 7.2° vs -14.7° ± 10.3°, p = 0.04), but there was no difference in postoperative tibial torsion between ATCs and STCs in this subgroup.
Conclusion:
The use of an anatomic tibial baseplate optimizes coverage by reducing posterolateral overhang and posteromedial underhang. It also achieved better rotational profiles compared to STCs. However, it resulted in a larger change in tibial torsion after TKA.
More Related Videos
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Bones of the Lower Limb: Femur and Patella
Bones of the Lower Limb: Tibia and Fibula
Muscles that Move the Leg
Anterior Compartment
The quadriceps femoris, the most visible muscle of the anterior compartment, is integral for leg extension and thigh flexion. It is formed by merging four distinct muscles — the vastus lateralis, vastus medialis, vastus intermedius, and rectus femoris. The quadriceps tendon, a shared tendon of the four quadriceps muscles, is affixed...
Ankle Joint