Automated detection of cardiac rest period for trigger delay calculation for image-based navigator coronary magnetic

Gregory Wood1,2, Alexandra Uglebjerg Pedersen3,4, Karl P Kunze5,6

  • 1Department of Cardiology, Aarhus University Hospital, Palle Juul Jensens Boulevard 99, 8200, Aarhus N, Denmark. gregory.wood@clin.au.dk.

Insights

Deep learning software automates trigger delay detection for coronary MRA, improving workflow efficiency. This method is faster than manual operator assessment, enhancing reproducibility in coronary artery disease imaging.

Area of Science:

  • Cardiovascular Imaging
  • Medical Imaging Analysis
  • Artificial Intelligence in Medicine

Background:

  • Coronary magnetic resonance angiography (coronary MRA) is a viable tool for diagnosing coronary artery disease (CAD).
  • Accurate trigger delay determination is crucial for minimizing cardiac motion artifacts in coronary MRA.
  • Current operator-dependent methods for trigger delay assessment can lead to inconsistent scan acquisition.

Purpose of the Study:

  • To evaluate the efficacy of deep learning (DL) software in automating cardiac rest period detection for coronary MRA.
  • To compare the performance of DL-based trigger delay determination against operator-led and formula-based methods.
  • To assess the impact of different trigger delay strategies on coronary MRA image quality and acquisition time.

Main Methods:

  • Thirty participants underwent motion-corrected coronary MRA using an iNAV-based sequence.
  • Trigger delay was determined using three methods: DL software, experienced operator, and a pre-existing formula.
  • Image quality was assessed by visible coronary vessel length and sharpness; acquisition times were recorded.

Main Results:

  • No significant differences in image quality (vessel length, sharpness) were observed across the three trigger delay determination methods.
  • The DL software method was significantly faster (106 ± 38.0 s) than the operator-led approach (168 ± 39.2 s) (p < 0.01).
  • The formula method showed slightly greater left circumflex vessel length compared to other methods.

Conclusions:

  • Deep learning software effectively automates optimal trigger delay detection for coronary MRA.
  • This automated approach can simplify workflow and enhance the reproducibility of coronary MRA acquisition.
  • DL-based trigger delay determination offers an efficient alternative to manual methods in CAD imaging.
Abstract