Moisture mitigation using a vented liner and a vented socket system for individuals with transfemoral amputation
Surya C Gnyawali1, Jeffrey A Denune2, Bryce Hockman2
1McGowan Institute for Regenerative Medicine (MIRM), Department of Surgery, University of Pittsburgh School of Medicine, Pittsburgh, PA, USA. scg92@pitt.edu.
A vented liner-socket system significantly reduces sweating and humidity for individuals with limb loss. This innovation improves comfort without compromising suspension, stability, or socket comfort.
Area of Science:
- Biomedical Engineering
- Prosthetics and Orthotics
- Human Physiology
Background:
- Heat and sweat buildup at the skin-liner interface is a significant challenge for amputees, leading to discomfort and potential skin breakdown.
- Current liners, often made of non-conductive materials, can exacerbate sweating during activity or warm conditions, affecting liner performance and residual limb health.
Purpose of the Study:
- To evaluate the effectiveness of a vented liner-socket system (VS) compared to a non-vented system (nVS) in reducing relative humidity (RH) at the skin-liner interface during physical activity.
- To assess the impact of the vented system on perceived sweating and user-reported outcomes related to prosthesis use.
Main Methods:
- Nine individuals with lower limb loss using a non-vented socket were randomized to either the vented liner-socket system (VS) or a non-vented socket (nVS).
- Participants engaged in a 20-minute treadmill walk to simulate activity.
- Relative humidity (RH) at the skin-liner interface was measured, and participants reported their perceived sweating levels and completed the Comprehensive Lower Limb Amputee Socket Survey (CLASS).
Main Results:
- The vented liner-socket system (VS) demonstrated a statistically significant reduction in relative humidity (RH) at the skin-liner interface (p=0.0002).
- Participants using the VS reported significantly improved perceived sweating (p=0.028).
- No significant differences were found in CLASS outcomes, indicating comparable suspension, stability, and comfort between the VS and nVS.
Conclusions:
- The vented liner-socket system effectively mitigates heat and humidity buildup at the skin-liner interface during activity.
- This system offers a tangible benefit for amputees experiencing sweating issues, enhancing comfort without negatively impacting prosthesis fit or function.
- The study provides objective evidence supporting the use of vented liner-socket systems for improved residual limb management in prosthetic users.
Related Concept Videos
Adaptations that Reduce Water Loss
Ostomy Care
An ostomy is a surgical procedure that creates an artificial opening from the intestines to the outside of the body, allowing for the rerouting of effluent. This opening is known as a stoma. A stoma usually protrudes above the skin surface, appearing pink or red, moist, and round, and it lacks nerve sensations.
There are different types of ostomies, including colostomies, ileostomies, and urostomies:
Cavity Drainage and Flashings in Masonry walls
Weep holes, strategically placed at the base of the cavity, are critical for draining accumulated water. These openings are created by leaving head...
Diabetic Foot Ulcer