EDOF intraocular lens design: shift in image plane vs object vergence
Pooria Omidi1, Alan Cayless2, Achim Langenbucher3
1Department of Experimental Ophthalmology, Saarland University, 66424, Homburg, Saarland, Germany. pooria.omidi@uks.eu.
Background:
To compare 2 different design scenarios of EDOF-IOLs inserted in the Liou-Brennan schematic model eye using raytracing simulation as a function of pupil size.
Methods:
Two EDOF IOL designs were created and optimized for the Liou-Brennan schematic model eye using Zemax ray tracing software. Each lens was optimized to achieve a maximum Strehl ratio for intermediate and far vision. In the first scenario, the object was located at infinity (O1), and the image plane was positioned at far focus (I1) and intermediate focus (I2) to emulate far and intermediate distance vision, respectively. In the second scenario, the image plane was fixed at I1 according to the first scenario. The object plane was set to infinity (O1) for far-distance vision and then shifted closer to the eye (O2) to reproduce the corresponding intermediate vision. The performance of both IOLs was simulated for the following 3 test conditions as a function of pupil size: a) O1 to I1, b) O1 to I2, and c) O2 to I1. To evaluate the imaging performance, we used the Strehl ratio, the root-mean-square (rms) of the spot radius, and the spherical aberration of the wavefront for various pupil sizes.
Results:
Evaluating the imaging performance of the IOLs shows that the imaging performance of the IOLs is essentially identical for object/image at O1/I1. Designed IOLs perform dissimilarly to each other in near-vision scenarios, and the simulations confirm that there is a slight difference in their optical performance.
Conclusion:
Our simulation study recommends considering the difference between object shift and image plane shift in design and test conditions to achieve more accurate pseudoaccommodation after cataract surgery.
More Related Videos
Related Concept Videos
Focusing of Light in the Eye
Depth Perception and Spatial Vision
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...