Related Experiment Video
Updated: Jul 15, 2025

Endoscopic Septoplasty with Limited Two-line Resection: Minimally Invasive Surgery for Septal Deviation
Published on: June 20, 2018
Screw Misposition and Detachment During Paranasal Augmentation: A Patient Report With Complication and Management
Panxi Yu1, Jianjian Lu1, Tong Wang2
1The Second Department of Craniomaxillofacial Surgery, Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College.
Abstract:
Paranasal concavity is esthetically displeasing and interferes with the overall outline of the face. Paranasal augmentation with alloplastic implants is a minimally invasive procedure with significant esthetic effects for paranasal concavity. When fixing the implants, the screws should be placed on the nasofrontal column, but are occasionally mispositioned on the anterior wall of the maxillary sinus. Here, the authors introduce a case of paranasal augmentation in which the self-tapping screw accidentally penetrated through the implant, broke through the anterior and medial walls of the maxillary sinus, and entered the nasal cavity. The screw was located by a computed tomography scan and was successfully removed under a nasal endoscope. The causes of this intraoperative complication were analyzed, emergent management was introduced, and a further preventive strategy was proposed.
More Related Videos
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
05:54Real-Time Dynamic Navigation System for the Precise Quad-Zygomatic Implant Placement in a Patient with a Severely Atrophic Maxilla
Published on: October 18, 2021