Harmonization of ANA testing challenge: quantification strategy to accurately predict end-point titers avoiding
Teresa Carbone1, Vito Pafundi2, Vito Ciardo3
1Immunopathology Laboratory, San Carlo Hospital, Potito Petrone St., 85100, Potenza, Italy.
Immunologic Research
|October 4, 2023
Summary
This study introduces a new method for antinuclear antibody (ANA) testing, using fluorescence intensity to predict antibody titers without serial dilutions. This approach significantly reduces lab time and improves result consistency.
Area of Science:
- Immunology
- Clinical Chemistry
- Laboratory Medicine
Background:
- Automated antinuclear antibody (ANA) analysis offers advantages, but predicting end-point titers without serial dilutions remains a challenge.
- Current methods for ANA titer determination often involve time-consuming serial dilutions, impacting laboratory efficiency.
Purpose of the Study:
- To develop and validate a conversion table correlating fluorescence signal intensity (FI) values with ANA end-point titers.
- To establish a method for predicting ANA titers from a single sample dilution, avoiding traditional serial dilutions.
Main Methods:
- Eight hundred ninety-four serum samples were analyzed for ANA using the Image Navigator System.
- Statistical discriminant analysis was employed to classify FI into ranges corresponding to conventional end-point titers.
- A validation study assessed agreement and error rates between visual readings and the conversion table for 1119 ANA-positive samples.
Main Results:
- A conversion table was computed, establishing FI ranges for end-point titers across different staining patterns.
- Overall agreement between the conversion table and visual readings was 98.4% for single-pattern samples (titers 1:160-1:2560).
- Concordance rates were high for nuclear (98.4%) and cytoplasmic (100%) patterns, with slightly lower rates for mitotic patterns (66.6%).
Conclusions:
- A novel quantification method for autoantibody titer assessment was developed, utilizing a single sample dilution.
- This method offers significant advantages for routine laboratories, including reduced hands-on time and improved result harmonization.
- The developed tool effectively predicts ANA end-point titers, enhancing efficiency in automated ANA analysis.
Keywords:
Antinuclear antibodyAutomated systemEnd-point titerFluorescence intensityIndirect immunofluorescenceMore Related Videos
Related Concept Videos
EDTA: Indirect and Alkalimetric Titration
949
Unlike direct titration, back-titration, and displacement titration, indirect titration is an EDTA titration method for quantifying anions. In the indirect titration method, anions are precipitated as their insoluble salts with excess metal ions. The filtrate containing the excess metal ions is directly titrated with standard EDTA until the endpoint is achieved. Another approach involves extracting the metal ion and back-titrating with standard EDTA to obtain the endpoint. In this way, the...
949
Titrimetric Methods: Types and Commonly Used Strategies
947
In chemistry, titrimetric methods are broadly classified into three types: volumetric, gravimetric, and coulometric. Volumetric titrations involve measuring the volume of a titrant of known concentration that is required to react completely with an analyte. In gravimetric titrations, the standard solution reacts with the analyte to form an insoluble precipitate, which is filtered, dried, and weighed. In coulometric titrations, current is applied to an electrochemical reaction until the reaction...