Aural microsuction: an analysis of post-procedure patient safety incidents
Niall J McInerney1,2, Nick O'Keeffe3,4, Tara Mackle3
1Department of Otolaryngology, Royal Victoria Eye and Ear Hospital, Dublin, Ireland. nialljmcinerney@rcsi.com.
Background:
Aural microsuction can be associated with delayed syncopal and vertiginous events, which can lead to serious adverse patient safety incidents.
Objective:
This article highlights the serious risks associated with microsuction.
Materials And Methods:
We performed a retrospective review of the hospital risk database over a 2-year period.
Results:
Three patients experienced severe adverse events post microsuction. Two patients fell down the stairs after microsuction, with one suffering a lower limb fracture. One patient experienced loss of consciousness and subsequent head trauma post microsuction.
Conclusions And Significance:
Whilst aural microsuction is universally accepted to be a low-risk procedure, subsequent serious morbidity can occur. Implementation and awareness of local safety policies are imperative to avoid post-procedure injury.
Insights
Aural microsuction, a common ear procedure, can lead to delayed dizziness and fainting, causing patient safety incidents. Awareness of these risks and safety policies is crucial to prevent injuries.
Area of Science:
- Otolaryngology
- Patient Safety
- Medical Procedures
Background:
- Aural microsuction, a widely used ear cleaning technique, is associated with potential delayed adverse events.
- These events, including syncope and vertigo, pose significant patient safety risks.
Purpose of the Study:
- To highlight the serious risks and potential adverse events linked to aural microsuction.
- To emphasize the importance of recognizing and mitigating these risks.
Main Methods:
- A retrospective review of a hospital risk database was conducted.
- Data were collected over a two-year period to identify adverse events post-microsuction.
Main Results:
- Three patients experienced severe adverse events following aural microsuction.
- Events included falls resulting in fractures and loss of consciousness with head trauma.
Conclusions:
- Despite being considered low-risk, aural microsuction can lead to significant patient morbidity.
- Implementing and adhering to local safety policies is essential for preventing post-procedure injuries.
Related Concept Videos
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Tracheostomy Suctioning II: Procedure
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Sputum Studies II: Culture and Sensitivity
Sputum culture and sensitivity is a medical procedure used to diagnose bacterial infections in the respiratory tract and select the most appropriate antibiotics for treatment. This process involves analyzing sputum samples of thick and opaque secretions produced in the lungs and airways. These samples are collected from patients and then sent to the laboratory for analysis.
The test can identify various pathogens responsible for respiratory infections, including Streptococcus,...