Dose difference between anisotropic analytical algorithm (AAA) and Acuros XB (AXB) caused by target's air content for
Takaaki Ito1,2, Hajime Monzen1, Kazuki Kubo1
1Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, Osakasayama, Osaka, Japan.
Background:
We clarified the dose difference between the anisotropic analytical algorithm (AAA) and Acuros XB (AXB) with increasing target's air content using a virtual phantom and clinical cases.
Materials And Methods:
Whole neck volumetric modulated arc therapy (VMAT) plan was transferred into a virtual phantom with a cylindrical air structure at the center. The diameter of the air structure was changed from 0 to 6 cm, and the target's air content defined as the air/planning target volume (PTV) in percent (air/PTV) was varied. VMAT plans were recalculated by AAA and AXB with the same monitor unit (MU) and multi-leaf collimator (MLC) motions. The dose at each air/PTV (5%-30%) was compared between each algorithm with D98%, D95%, D50% and D2% for the PTV. In addition, MUs were also compared with the same MLC motions between the D95% prescription with AAA (AAA_D95%), AXB_D95%, and the prescription to 100% minus air/PTV (AXB_D100%-air/PTV) in clinical cases of head and neck (HNC).
Results:
When air/PTV increased (5-30%), the dose differences between AAA and AXB for D98%, D95%, D50% and D2% were 3.08-15.72%, 2.35-13.92%, 0.63-4.59%, and 0.14-6.44%, respectively. At clinical cases with air/PTV of 5.61% and 28.19%, compared to AAA_D95%, the MUs differences were, respectively, 2.03% and 6.74% for AXB_D95% and 1.80% and 0.50% for AXB_D100%-air/PTV.
Conclusion:
The dose difference between AAA and AXB increased as the target's air content increased, and AXB_D95% resulted in a dose escalation over AAA_D95% when the target's air content was ≥ 5%. The D100%-air/PTV of PTV using AXB was comparable to the D95% of PTV using AAA.
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...