Related Experiment Video
Updated: Jul 14, 2025

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Comparing lung aeration and respiratory effort using two different spontaneous breathing trial: T-piece vs pressure
Raquel Bosch-Compte1, Francisco José Parrilla1, Rosana Muñoz-Bermúdez1
1Critical Care Department, Hospital del Mar, Critical Illness Research Group (GREPAC), Institut Hospital del Mar d'Investigacions Mèdiques (IMIM), Barcelona, Spain.
Objective:
To assess the changes in lung aeration and respiratory effort generated by two different spontaneous breathing trial (SBT): T-piece (T-T) vs pressure support ventilation (PSV).
Design:
Prospective, interventionist and randomized study.
Setting:
Intensive Care Unit (ICU) of Hospital del Mar.
Participants:
Forty-three ventilated patients for at least 24 h and considered eligible for an SBT were included in the study between October 2017 and March 2020.
Interventions:
30-min SBT with T-piece (T-T group, 20 patients) or 8-cmH2O PSV and 5-cmH2O positive end expiratory pressure (PSV group, 23 patients).
Main Variables Of Interest:
Demographics, clinical data, physiological variables, lung aeration evaluated with electrical impedance tomography (EIT) and lung ultrasound (LUS), and respiratory effort using diaphragmatic ultrasonography (DU) were collected at different timepoints: basal (BSL), end of SBT (EoSBT) and one hour after extubation (OTE).
Results:
There were a loss of aeration measured with EIT and LUS in the different study timepoints, without statistical differences from BSL to OTE, between T-T and PSV [LUS: 3 (1, 5.5) AU vs 2 (1, 3) AU; p = 0.088; EELI: -2516.41 (-5871.88, 1090.46) AU vs -1992.4 (-3458.76, -5.07) AU; p = 0.918]. Percentage of variation between BSL and OTE, was greater when LUS was used compared to EIT (68.1% vs 4.9%, p ≤ 0.001). Diaphragmatic excursion trend to decrease coinciding with a loss of aeration during extubation.
Conclusion:
T-T and PSV as different SBT strategies in ventilated patients do not show differences in aeration loss, nor estimated respiratory effort or tidal volume measured by EIT, LUS and DU.
Related Concept Videos
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.