1Department of Cardiac Surgery, Asklepios Klinik St Georg, Hamburg, Germany.
Objectives:
Patients on ticagrelor undergoing urgent cardiac surgery are at high risk for perioperative bleeding complications. We sought to determine whether intraoperative hemoadsorption could remove ticagrelor and lower circulating drug concentrations.
Methods:
The hemoadsorption device was incorporated in the cardiopulmonary bypass (CPB) circuit and remained active for the duration of the pump run. Blood samples were collected before and after CPB. The main objective of the current analysis was to compare mean total plasma ticagrelor levels (ng/mL) at baseline with ticagrelor levels obtained at the end of CPB. Plasma ticagrelor levels were measured at a certified outside laboratory (LabConnect). Data are presented as mean ± standard deviation.
Results:
A total of 11 patients undergoing urgent coronary artery bypass grafting at 3 institutions were included (mean age, 67.9 ± 9.9 years; 91% male; mean European System for Cardiac Operative Risk Evaluation II of 3.0 ± 3.3%; range, 0.7%-12.4%). Mean intraoperative hemoadsorption duration was 97.1 ± 43.4 minutes with a mean flow rate through the device of 422.9 ± 40.3 mL/min. Mean ticagrelor levels pre-CPB were 103.5 ± 63.8 ng/mL compared with mean post-CPB levels of 34.0 ± 17.5 ng/mL, representing a significant 67.1% reduction (P < .001). Intraoperative integration of the device was simple and safe without any device-related adverse events reported.
Conclusions:
This is the first in vivo report demonstrating that intraoperative hemoadsorption can efficiently remove ticagrelor and significantly reduce circulating drug levels. Whether active ticagrelor removal can reduce serious perioperative bleeding in patients undergoing urgent cardiac surgery is currently being evaluated in the double-blinded, randomized Safe and Timely Antithrombotic Removal-Ticagrelor (STAR-T) trial.